DMS-40PC Series
Actual Size
4½ Digit, LED Display
Precision, Miniature
Digital Panel Voltmeters
Features
Precision, autozeroing, factory-
calibrated A/D converter
Scientific-grade accuracy, ±2 counts
Miniature size:
2.17" x 0.92" x 0.56"
55mm x 23mm x 14mm
Large (0.52"/13.2mm) LED display
Choice of red, green or yellow colors
High-intensity or low-power red
LED's optional
Single +5V supply (175mW for low-
power models)
Epoxy-encapsulated, 12-pin DIP with
built-in color filter and bezel
3 differential input voltage ranges
DISPLAY HOLD and TEST functions
Optional BCD data outputs for CPU
interface
0 to +50°C temperature range
DMS-40PC Series, 4½ Digit, LED Display, Miniature DPM's are fully self-contained,
component-like, plug-in meters that provide scientific-grade accuracy (typically ±2
counts or ±0.005% of full scale) and outstanding reliability at a very affordable price.
Within its miniature (2.17" x 0.92" x 0.56"), epoxy-encapsulated package, each
meter contains a precision reference circuit; a high-resolution, autozeroing, factory-
calibrated A/D converter; and a large (0.52"/13.2mm), easy-to-read, LED display.
LED's are available in red, yellow and green colors. Red LED's are also offered in
high-intensity or low-power versions.
The versatile design of the DMS-40PC Series assures trouble-free installation and
long-term operation. Differential input voltage ranges include ±2V, ±20V and ±200V.
Input impedance is a minimum 800k. Non-inverting inputs are overvoltage protected
to ±250V, and CMRR is typically 86dB (dc to 60Hz).
The DMS-40PC's epoxy-encapsulated package has an integral bezel and color
filter. The moisture and vibration-proof package is extremely rugged and well suited for
harsh environments and extended temperatures. Devices are fully specified for
0 to +50°C operation.
All models operate from a single +5V supply and typically consume 500mW. Low-
power models, whose display is just as bright as standard models, typically consume
175mW. DISPLAY TEST and HOLD functions are standard on each meter, and a
complete set of BCD outputs are optional for sending data to CPU's or remote
displays.
1
+5V SUPPLY
2
DISPLAY TEST
3
5V RETURN
4
DP 3
5
DP 2
6
DP 1
+5V
V+
DC/DC
CONVERTER
–5V
V–
0 Vdc
A/D
CONVERTER
DATA
0.01µF
1
R2
12
(–) INPUT LO
R1
909k
11
(+) INPUT HI
10 ANALOG
COMMON
2.5V
REF.
9
DISPLAY HOLD
8
DP4
+5V
7
REFERENCE IN/OUT
1 R2 is not used on ±2V (-1) models.
R2 = 101k on ±20V (-2) models and 9.2k on ±200V (-3) models.
Figure 1. DMS-40PC Series Simplified Schematic
DATEL, Inc., Mansfield, MA 02048 (USA) Tel: (508)339-3000, (800)233-2765 Fax: (508)339-6356 Email: [email protected] Internet: www.datel.com

1 page




DMS-40PC pdf
4½ D I G I T, L E D D I S P L AY D I G I TA L PA N E L V O LT M E T E R S
DMS-40PC
Applications
5. Engineering Scaling and Input Attenuation: In many
applications, the input signal must be scaled, or divided down,
before being applied to the DMS-40PC. In some situations, the
input signal exceeds the full-scale range of the meter, and in
other applications, a direct one-to-one relation does not exist
between the input voltage and the desired display reading. For an
example of the second situation, assume an input voltage of
1.0Vdc is required to display a reading of "7500" when applied to
a ±2V input meter. An input divider circuit, constructed with two
series resistors with an output-to-input ratio of 0.75, will scale the
1.0V signal down to 0.75V.
Engineering scaling and input attenuation are discussed in Ap
Note 4 of the DATEL Digital Panel Voltmeter Catalog. Compen-
sating for inputs which have a zero offset is also discussed.
6. BCD Outputs (DMS-40PC-X-RS and -RL-BCD Models Only):
Models with a "-BCD" suffix have 12 extra output pins, labeled
A through L (6 per side), that provide the following information:
multiplexed serial BCD data, digit drive, and polarity indication.
The data present on pins A - L simultaneously drives the meter's
internal LED display. All functions on pins A - L are 5V CMOS
compatible, however, they are only rated to drive one, 74LS
series, TTL load. CMOS logic IC's, for example 74HC or 74HCT
series devices, should be used when more fan-out capability
is required.
As the timing diagram in Figure 5 indicates, 100Hz is the
optimum display scan rate when the BCD outputs are used to
drive external LED displays. Faster scan rates, while permissible,
are not necessary. Slower scanning, however, may result in
noticeable "flickering" of the display. Common anode LED's,
combined with a 74LS247 BCD to seven-segment decoder, is the
simplest way of implementing an external display. A seventy-five
to one-hundred Ohm resistor on each LS247 segment-drive
output provides adequate display brightness.
The functions of pins A - L are listed below:
BCD DATA (pins I - L): Four lines are used for BCD (Binary Coded
Decimal) data outputs, representing the numbers 0 - 9. Positive-logic
convention (a high represents a "1") is used.
DIGIT DRIVE ( pins A - E): These five outputs, when gated with
STROBE, can be used to direct the BCD DATA into external latches.
DIGIT DRIVE outputs may also be connected directly to the bases of
NPN transistors in remote-display configurations. The digits are
scanned right-to-left, i.e., MSD (DIGIT 1 DRIVE) to LSD (DIGIT 5
DRIVE).
BUSY (pin G): This is a status pin that goes high at the start of an
analog-to-digital (A/D) conversion cycle and remains high until the
conversion ends.
STROBE (pin H): A string of five sequential, active-low, STROBE
pulses are output (after BUSY goes low) indicating the end of a
conversion and the availability of new data. The 5 STROBE pulses
occur only once per conversion, or 2.5 times each second. STROBE
can be used to latch the BCD data (on pins I - L) into external
latches. Either edge can be used since STROBE (~ 5µsec wide) is
active only in the center of the corresponding DIGIT DRIVE and BCD
DATA outputs.
POLARITY (pin F): This pin, which is also used inside the DMS-
40PC to drive the negative-sign segment, indicates whether the last
input signal conversion was positive (POLARITY set high) or
negative (POLARITY set low). POLARITY, unlike BCD DATA, is not
multiplexed. For data latching purposes, POLARITY should be
sampled during the STROBE pulse for digit 1 (most significant digit).
DIGIT DRIVE 1
(MSD)
DIGIT DRIVE 2
DIGIT DRIVE 3
DIGIT DRIVE 4
DIGIT DRIVE 5
(LSD)
BUSY
STROBE OUT
BCD DATA
(4 LINES)
Approx. 2msec
Approx. 2msec
Approx. 8msec
Approx. 5µsec
Data from
Previous Conversion
5 Strobes Only Per Conversion
1 234 51 23 4 5
MSD
LSD
Figure 6. BCD Data Timing Diagram
5

5 Page










PáginasTotal 6 Páginas
PDF Descargar[ Datasheet DMS-40PC.PDF ]




Hoja de datos destacado

Número de piezaDescripciónFabricantes
DMS-40PCLED Display Precision / Miniature Digital Panel VoltmetersDatel
Datel

Número de piezaDescripciónFabricantes
SLA6805M

High Voltage 3 phase Motor Driver IC.

Sanken
Sanken
SDC1742

12- and 14-Bit Hybrid Synchro / Resolver-to-Digital Converters.

Analog Devices
Analog Devices


DataSheet.es es una pagina web que funciona como un repositorio de manuales o hoja de datos de muchos de los productos más populares,
permitiéndote verlos en linea o descargarlos en PDF.


DataSheet.es    |   2020   |  Privacy Policy  |  Contacto  |  Buscar